PRODUCT DETAILS
Features
Lightweight handheld meter with option of eStation docking system
Dried-reagent strips stored at room temperature for true ready-to-use
Excellent correlation to central laboratory results
Versatile sample types: venous whole blood,citrate whole blood
Smallest blood sample volume
Stable external QC controls
Flexible reference intervals set by operator
Built-in barcode scanner for accuracy and efficiency
Built-in communication port for printers or computer networks
Patented RACE TM technology for rapid and accurate PT and APTT determination
12 months warranty
Specification
Animal Species
Canine (dog)
Feline (cat)
Sample Type
Fresh venous whole blood
Citrate whole blood
Sample Volume
10 µL
Time to Results
< 7 minutes
Reportable Time Range
PT: 10 to 80 seconds
APTT: 10 to 420 seconds
Reference Intervals
Canine (dog)
-PT: 14 to 19 seconds
-APTT: 75 to 105 seconds
Feline (cat)
-PT: 15 to 21 seconds
-APTT: 94 to 125 seconds
Precision
In-house lot-to-lot and
day-to-day precision
<5.0 CV% (mean value)
External QC Control
Lyophilized whole blood
preparations
Level I, Level II
Operating Temperature Range
10°C to 35°C
Operating Humidity Range
10% to 90% RH
Shelf Life at Room Temperature
12 months at 15°C to 30°C